Recall Z-1811-2015— CareFusion 303, Inc.

Class II · Terminated

Class II FDA medical device recall by CareFusion 303, Inc., initiated 2015-05-13, terminated 2016-05-20. It names one FDA 510(k) clearance: K051641. Product: Alaris PC unit model 8015, Infusion pump. software version 9.17 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The software is embedded in the Alaris PC unit.. Reason: An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication..

Recalling firm
CareFusion 303, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-05-13
Terminated
2016-05-20
Firm location
San Diego, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Alaris PC unit model 8015, Infusion pump. software version 9.17 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The software is embedded in the Alaris PC unit.

Reason for recall

An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.