Recall Z-1806-2022— Mckesson Medical-Surgical Inc. Corporate Office
Class III · Ongoing
Class III FDA medical device recall by Mckesson Medical-Surgical Inc. Corporate Office, initiated 2022-05-25. It names 2 FDA 510(k) clearances: K183645, K090453. Product: a. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WOUND GELLING FIBERHIGH-ABSRB 4"X4 3/4" (10/BX) Catalog #87400. b. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WND GELLING FIBER HIABSRB AG 4X4.75 (10/BX 10BX/CS Catalog #177400. c. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WOUND GELLING FIBERHI-ABSRB AG 2X2 (10/BX 10BX/CS Catalog # 177200. Reason: Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences..
- Recalling firm
- Mckesson Medical-Surgical Inc. Corporate Office
- Classification
- Class III
- Status
- Ongoing
- Initiated
- 2022-05-25
- Firm location
- Richmond, VA, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
a. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WOUND GELLING FIBERHIGH-ABSRB 4"X4 3/4" (10/BX) Catalog #87400. b. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WND GELLING FIBER HIABSRB AG 4X4.75 (10/BX 10BX/CS Catalog #177400. c. ADVANCED MEDICAL SOLUTIONS McKesson DRESSING, WOUND GELLING FIBERHI-ABSRB AG 2X2 (10/BX 10BX/CS Catalog # 177200
Reason for recall
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.