Recall Z-1805-2014— Ortho-Clinical Diagnostics
Class II · Terminated
Class II FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2014-05-12, terminated 2017-03-23. It names 2 FDA 510(k) clearances: K063164, K041799. Product: VITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6801873 -- For in vitro diagnostic use only --- VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate the VITROS 5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.. Reason: Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28..
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-05-12
- Terminated
- 2017-03-23
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
2 on FDA's record- K063164 — VITROS CHEMISTRY PRODUCTS THC REAGENT, CALIBRATOR KIT 30, FS CALIBRATOR 1, AND DAT PERFORMANCE VERIFIERS I, II, III, IV
- K041799 — VITROS CHEMISTRY PRODUCTS HSCRP REAGENT, VITROS CHEMISTRY PRODUCTS CALIBRATOR KIT 17, VITROS CHEMISTRY PRODUCTS FS
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6801873 -- For in vitro diagnostic use only --- VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate the VITROS 5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Reason for recall
Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.