Recall Z-1801-2023— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Class I · Ongoing

Class I FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2023-05-10. FDA's recall record names no specific 510(k) clearance for this event. Product: ICD-VR DVEX2E4 EV ICD OUS EV4, Model Number DVEX2E4; Implantable Cardioverter Defibrillators. Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds..

Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Classification
Class I
Status
Ongoing
Initiated
2023-05-10
Firm location
Mounds View, MN, United States

Product description

ICD-VR DVEX2E4 EV ICD OUS EV4, Model Number DVEX2E4; Implantable Cardioverter Defibrillators

Reason for recall

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Data sourced from openFDA. This site is unofficial and independent of the FDA.