Recall Z-1800-2020— Bio-Rad Labs

Class II · Completed

Class II FDA medical device recall by Bio-Rad Labs, initiated 2019-04-03. It names one FDA 510(k) clearance: K042644. Product: Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260; Lot number 18200A - Product Usage: The assay is an aid in the diagnosis of celiac disease (gluten-sensitive enteropathy) and should be used in conjunction with other serological tests and clinical findings.. Reason: An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged..

Recalling firm
Bio-Rad Labs
Classification
Class II
Status
Completed
Initiated
2019-04-03
Firm location
Benicia, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260; Lot number 18200A - Product Usage: The assay is an aid in the diagnosis of celiac disease (gluten-sensitive enteropathy) and should be used in conjunction with other serological tests and clinical findings.

Reason for recall

An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged.