Recall Z-1796-2020— Merit Medical Systems, Inc.
Class II · Terminated
Class II FDA medical device recall by Merit Medical Systems, Inc., initiated 2020-03-11, terminated 2022-03-02. It names one FDA 510(k) clearance: K180723. Product: PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT. Reason: During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon..
- Recalling firm
- Merit Medical Systems, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-03-11
- Terminated
- 2022-03-02
- Firm location
- South Jordan, UT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT
Reason for recall
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.