Recall Z-1796-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-12-14, terminated 2020-12-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase). Reason: Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-to-lot limits..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-12-14
Terminated
2020-12-03
Firm location
Newark, DE, United States

Product description

Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase)

Reason for recall

Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-to-lot limits.

Data sourced from openFDA. This site is unofficial and independent of the FDA.