Recall Z-1795-2021— Philips North America Llc
Class II · Ongoing
Class II FDA medical device recall by Philips North America Llc, initiated 2021-04-23. It names one FDA 510(k) clearance: K181479. Product: Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109 - Ingenia Ambition X. Reason: The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet..
- Recalling firm
- Philips North America Llc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-04-23
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109 - Ingenia Ambition X
Reason for recall
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.