Recall Z-1792-2014— Roche Diagnostics Operations, Inc.
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2014-05-21, terminated 2015-06-04. It names one FDA 510(k) clearance: K060373. Product: cobas 8000 analyzer series Operators Manual Roche cobas Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N) Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.. Reason: Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms..
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-05-21
- Terminated
- 2015-06-04
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
cobas 8000 analyzer series Operators Manual Roche cobas Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N) Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Reason for recall
Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
Data sourced from openFDA. This site is unofficial and independent of the FDA.