Recall Z-1789-2021— Terumo Cardiovascular Systems Corporation
Class II · Terminated
Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2021-04-16, terminated 2023-11-29. It names one FDA 510(k) clearance: K071521. Product: The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right). The system has the capacity to circulate water at a rate of up to 6.5 gal./min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Device Name / Model Number: HX2 Temperature Management System (P/N 809810) Catalog Number: 809810. Reason: Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices..
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-04-16
- Terminated
- 2023-11-29
- Firm location
- Ann Arbor, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right). The system has the capacity to circulate water at a rate of up to 6.5 gal./min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Device Name / Model Number: HX2 Temperature Management System (P/N 809810) Catalog Number: 809810
Reason for recall
Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.
Data sourced from openFDA. This site is unofficial and independent of the FDA.