Recall Z-1786-2021— Avid Medical, Inc.

Class I · Terminated

Class I FDA medical device recall by Avid Medical, Inc., initiated 2021-04-09, terminated 2022-07-18. FDA's recall record names no specific 510(k) clearance for this event. Product: FISTULA (ON/OFF) ALL IN ONE BAG. Medical convenience kit, Avid Medical Part Number: VAMK024-03. Reason: Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus penicillioides, which may result in a breach of the package integrity. A breach of package integrity may contaminate other components in AVID Medical kits..

Recalling firm
Avid Medical, Inc.
Classification
Class I
Status
Terminated
Initiated
2021-04-09
Terminated
2022-07-18
Firm location
Toano, VA, United States

Product description

FISTULA (ON/OFF) ALL IN ONE BAG. Medical convenience kit, Avid Medical Part Number: VAMK024-03

Reason for recall

Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus penicillioides, which may result in a breach of the package integrity. A breach of package integrity may contaminate other components in AVID Medical kits.

Data sourced from openFDA. This site is unofficial and independent of the FDA.