Recall Z-1786-2013— Steris Corporation

Class II · Terminated

Class II FDA medical device recall by Steris Corporation, initiated 2013-05-20, terminated 2013-08-28. FDA's recall record names no specific 510(k) clearance for this event. Product: STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.. Reason: A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati.

Recalling firm
Steris Corporation
Classification
Class II
Status
Terminated
Initiated
2013-05-20
Terminated
2013-08-28
Firm location
Mentor, OH, United States

Product description

STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.

Reason for recall

A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati

Data sourced from openFDA. This site is unofficial and independent of the FDA.