Recall Z-1785-2022— CareFusion 303, Inc.

Class II · Ongoing

Class II FDA medical device recall by CareFusion 303, Inc., initiated 2022-05-24. FDA's recall record names no specific 510(k) clearance for this event. Product: BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex operating with Windows 7 . MedBank 200 Main operating with Windows 7 . MedBank 500 Main operating with Windows 7. An automated dispensing cabinet intended to securely store and dispense medications to health care providers in a clinical setting.. Reason: The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities..

Recalling firm
CareFusion 303, Inc.
Classification
Class II
Status
Ongoing
Initiated
2022-05-24
Firm location
San Diego, CA, United States

Product description

BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex operating with Windows 7 . MedBank 200 Main operating with Windows 7 . MedBank 500 Main operating with Windows 7. An automated dispensing cabinet intended to securely store and dispense medications to health care providers in a clinical setting.

Reason for recall

The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities.

Data sourced from openFDA. This site is unofficial and independent of the FDA.