Recall Z-1780-2017— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2017-03-10, terminated 2018-03-12. It names one FDA 510(k) clearance: K150087. Product: Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD.. Reason: GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2017-03-10
Terminated
2018-03-12
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Vivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD.

Reason for recall

GE Healthcare has become aware of an issue where a patient other than the intended is incorrectly selected by the operator in situations where DICOM Worklist search response time is slow. This issue is limited to certain Vivid Ultrasound systems and can result in the incorrect patient information showing on the screen during the exam. If not detected by the operator, images from the actual (intended) patient will be stored under the incorrect patient after the exam. This issue could lead to misdiagnosis. There have been no injuries reported because of this issue.

Data sourced from openFDA. This site is unofficial and independent of the FDA.