Recall Z-1778-2019— Maquet Cardiovascular Us Sales, Llc
Class II · Terminated
Class II FDA medical device recall by Maquet Cardiovascular Us Sales, Llc, initiated 2019-05-20, terminated 2023-12-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.. Reason: Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process..
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-05-20
- Terminated
- 2023-12-15
- Firm location
- Wayne, NJ, United States
Product description
Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Reason for recall
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Data sourced from openFDA. This site is unofficial and independent of the FDA.