Recall Z-1776-2019— Conformis, Inc.
Class II · Terminated
Class II FDA medical device recall by Conformis, Inc., initiated 2019-03-29, terminated 2020-04-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Conformis iTotal Hip Replacement System: HBS-033-00xx-020101, (HIP-PATIENT SPECIFIC STEM RIGHT). Reason: Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants..
- Recalling firm
- Conformis, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-03-29
- Terminated
- 2020-04-28
- Firm location
- Billerica, MA, United States
Product description
Conformis iTotal Hip Replacement System: HBS-033-00xx-020101, (HIP-PATIENT SPECIFIC STEM RIGHT)
Reason for recall
Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.