Recall Z-1775-2025— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Class II · Ongoing
Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2025-03-24. It names one FDA 510(k) clearance: K193215. Product: Intera 1.5T Explorer/Nova Dual, Model Number: 781108;. Reason: The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury..
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-03-24
- Firm location
- Best, Netherlands
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Intera 1.5T Explorer/Nova Dual, Model Number: 781108;
Reason for recall
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Data sourced from openFDA. This site is unofficial and independent of the FDA.