Recall Z-1775-2021— Fujifilm Medical Systems U.S.A., Inc.
Class II · Terminated
Class II FDA medical device recall by Fujifilm Medical Systems U.S.A., Inc., initiated 2021-04-30, terminated 2024-07-29. It names one FDA 510(k) clearance: K151859. Product: Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system. Reason: Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence.
- Recalling firm
- Fujifilm Medical Systems U.S.A., Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-04-30
- Terminated
- 2024-07-29
- Firm location
- Lexington, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system
Reason for recall
Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence