Recall Z-1772-2022— Randox Laboratories Ltd.

Class II · Terminated

Class II FDA medical device recall by Randox Laboratories Ltd., initiated 2022-08-01, terminated 2024-12-13. FDA's recall record names no specific 510(k) clearance for this event. Product: The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.. Reason: ***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg.

Recalling firm
Randox Laboratories Ltd.
Classification
Class II
Status
Terminated
Initiated
2022-08-01
Terminated
2024-12-13
Firm location
Crumlin (North), N/A, Ireland

Product description

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Reason for recall

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

Data sourced from openFDA. This site is unofficial and independent of the FDA.