Class II FDA medical device recall by Medtronic Minimed, initiated 2021-03-15. FDA's recall record names no specific 510(k) clearance for this event. Product: MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ PUMP MMT-1740KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KFX 670G V6.3 SF MG CLIN FR;¿ PUMP MMT-1741KFX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KX 670G V6.3 SF MG CLIN GER;¿ PUMP MMT-1741KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KXS 670G V6.3 SF MG CLIN IS;¿ PUMP MMT-1741KXS 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1742KX 670G V6.1 MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN SLO;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 MM;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 EU MM;¿. Reason: Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery..

Recalling firm
Medtronic Minimed
Classification
Class II
Status
Ongoing
Initiated
2021-03-15
Firm location
Northridge, CA, United States

Product description

MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ PUMP MMT-1740KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KFX 670G V6.3 SF MG CLIN FR;¿ PUMP MMT-1741KFX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KX 670G V6.3 SF MG CLIN GER;¿ PUMP MMT-1741KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KXS 670G V6.3 SF MG CLIN IS;¿ PUMP MMT-1741KXS 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1742KX 670G V6.1 MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN SLO;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 MM;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 EU MM;¿

Reason for recall

Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.

Data sourced from openFDA. This site is unofficial and independent of the FDA.