Recall Z-1771-2019— Philips North America, LLC

Class III · Terminated

Class III FDA medical device recall by Philips North America, LLC, initiated 2019-04-29, terminated 2024-11-22. It names one FDA 510(k) clearance: K153702. Product: Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information Center iX B.01 upgrade (or later) Product Usage: The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors. Reason: Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values.

Recalling firm
Philips North America, LLC
Classification
Class III
Status
Terminated
Initiated
2019-04-29
Terminated
2024-11-22
Firm location
Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips Patient Information Center iX configured with a regional setting which uses a comma for a decimal separator: 866389 Patient Information Center iX B.01 (or later); 866390 Patient Information Center iX B.01 upgrade (or later) Product Usage: The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors

Reason for recall

Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values

Data sourced from openFDA. This site is unofficial and independent of the FDA.