Recall Z-1765-2020— Siemens Medical Solutions USA, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2020-04-06, terminated 2020-10-27. It names one FDA 510(k) clearance: K061054. Product: System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions.. Reason: It was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total..
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-04-06
- Terminated
- 2020-10-27
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions.
Reason for recall
It was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.