Recall Z-1755-2024— Johnson & Johnson Surgical Vision, Inc.

Class II · Ongoing

Class II FDA medical device recall by Johnson & Johnson Surgical Vision, Inc., initiated 2024-03-14. It names one FDA 510(k) clearance: K121721. Product: Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.. Reason: Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms..

Recalling firm
Johnson & Johnson Surgical Vision, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-03-14
Firm location
Irvine, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

Reason for recall

Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.