Recall Z-1755-2023— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class I · Ongoing
Class I FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2023-05-10. FDA's recall record names no specific 510(k) clearance for this event. Product: CRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators. Reason: There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds..
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-05-10
- Firm location
- Mounds View, MN, United States
Product description
CRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Data sourced from openFDA. This site is unofficial and independent of the FDA.