Recall Z-1747-2017— Siemens Healthcare Diagnostics, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2017-03-08, terminated 2020-01-22. It names one FDA 510(k) clearance: K982680. Product: Siemens ADVIA Centaur BR assay Cat No. 02419937, SMN 10333349 (1-pack/50 tests) BR Assay for CA 27.29 The ADVIA Centaur¿ BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur¿, ADVIA Centaur¿ XP, and ADVIA Centaur¿ XPT systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.. Reason: Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems.

Recalling firm
Siemens Healthcare Diagnostics, Inc
Classification
Class II
Status
Terminated
Initiated
2017-03-08
Terminated
2020-01-22
Firm location
East Walpole, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Siemens ADVIA Centaur BR assay Cat No. 02419937, SMN 10333349 (1-pack/50 tests) BR Assay for CA 27.29 The ADVIA Centaur¿ BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur¿, ADVIA Centaur¿ XP, and ADVIA Centaur¿ XPT systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.

Reason for recall

Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems