Recall Z-1746-2022— Qiagen Sciences LLC

Class II · Ongoing

Class II FDA medical device recall by Qiagen Sciences LLC, initiated 2021-11-21. FDA's recall record names no specific 510(k) clearance for this event. Product: QIAcube Connect MDx, Model No. 9003070. Reason: During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results..

Recalling firm
Qiagen Sciences LLC
Classification
Class II
Status
Ongoing
Initiated
2021-11-21
Firm location
Germantown, MD, United States

Product description

QIAcube Connect MDx, Model No. 9003070

Reason for recall

During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.