Recall Z-1744-2026— Medtronic MiniMed, Inc.

Class II · Ongoing

Class II FDA medical device recall by Medtronic MiniMed, Inc., initiated 2026-02-13. FDA's recall record names no specific 510(k) clearance for this event. Product: MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851). Reason: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death..

Recalling firm
Medtronic MiniMed, Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-02-13
Firm location
Northridge, CA, United States

Product description

MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

Reason for recall

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Data sourced from openFDA. This site is unofficial and independent of the FDA.