Recall Z-1742-2013— Stryker Instruments Div. of Stryker Corporation

Class II · Terminated

Class II FDA medical device recall by Stryker Instruments Div. of Stryker Corporation, initiated 2013-07-01, terminated 2013-09-16. It names one FDA 510(k) clearance: K040300. Product: RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This device can also be used with the Total Performance System (TPSTM).. Reason: The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100.

Recalling firm
Stryker Instruments Div. of Stryker Corporation
Classification
Class II
Status
Terminated
Initiated
2013-07-01
Terminated
2013-09-16
Firm location
Portage, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This device can also be used with the Total Performance System (TPSTM).

Reason for recall

The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100

Data sourced from openFDA. This site is unofficial and independent of the FDA.