Recall Z-1741-2015— Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Biomet, Inc., initiated 2015-05-01, terminated 2015-12-08. It names one FDA 510(k) clearance: K111376. Product: Unilateral external fixation ankle clamp Product Usage: Unilateral external fixation for use treatment of bone conditions amenable to treatment by use of external fixation modality. Reason: Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately reconnect the fixator to the bone..
- Recalling firm
- Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-05-01
- Terminated
- 2015-12-08
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Unilateral external fixation ankle clamp Product Usage: Unilateral external fixation for use treatment of bone conditions amenable to treatment by use of external fixation modality
Reason for recall
Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately reconnect the fixator to the bone.