Recall Z-1733-2015— Roche Diabetes Care, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diabetes Care, Inc., initiated 2015-04-22, terminated 2016-11-03. It names one FDA 510(k) clearance: K111353. Product: AccuChek Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System.. Reason: Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u.

Recalling firm
Roche Diabetes Care, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-04-22
Terminated
2016-11-03
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AccuChek Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System.

Reason for recall

Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u