Recall Z-1733-2013— Terumo Cardiovascular Systems Corporation
Class II · Terminated
Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2013-02-27, terminated 2016-02-17. It names one FDA 510(k) clearance: K972962. Product: CDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.. Reason: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the.
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-02-27
- Terminated
- 2016-02-17
- Firm location
- Ann Arbor, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
Reason for recall
Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the