Recall Z-1731-2022— Exactech, Inc.
Class II · Ongoing
Class II FDA medical device recall by Exactech, Inc., initiated 2022-08-11. It names one FDA 510(k) clearance: K100269. Product: GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. GXL NEUTRAL LINER, G0 28MM ID, 130-28-50; b. GXL NEUTRAL LINER, G00 22MM ID, 130-22-70; c. NOVATION GXL LIPPED LINER, G0 28MM ID, 132-28-50; d. NOVATION GXL LIPPED LINER, G00 22MM ID, 132-22-70;. Reason: Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear..
- Recalling firm
- Exactech, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-08-11
- Firm location
- Gainesville, FL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. GXL NEUTRAL LINER, G0 28MM ID, 130-28-50; b. GXL NEUTRAL LINER, G00 22MM ID, 130-22-70; c. NOVATION GXL LIPPED LINER, G0 28MM ID, 132-28-50; d. NOVATION GXL LIPPED LINER, G00 22MM ID, 132-22-70;
Reason for recall
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.