Recall Z-1726-2014— Leica Microsystems, Inc.
Class II · Terminated
Class II FDA medical device recall by Leica Microsystems, Inc., initiated 2013-09-11, terminated 2017-04-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.. Reason: The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches)..
- Recalling firm
- Leica Microsystems, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-09-11
- Terminated
- 2017-04-24
- Firm location
- Buffalo Grove, IL, United States
Product description
Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.
Reason for recall
The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).