Recall Z-1725-2014— Leica Microsystems, Inc.

Class II · Terminated

Class II FDA medical device recall by Leica Microsystems, Inc., initiated 2013-09-11, terminated 2017-04-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.. Reason: The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches)..

Recalling firm
Leica Microsystems, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-09-11
Terminated
2017-04-24
Firm location
Buffalo Grove, IL, United States

Product description

Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Reason for recall

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).