Recall Z-1723-2024— Zimmer, Inc.
Class II · Ongoing
Class II FDA medical device recall by Zimmer, Inc., initiated 2024-03-20. It names one FDA 510(k) clearance: K070906. Product: ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-017-06. Reason: Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention.
- Recalling firm
- Zimmer, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-03-20
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-017-06
Reason for recall
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
Data sourced from openFDA. This site is unofficial and independent of the FDA.