Recall Z-1721-2025— Getinge Disinfection Ab

Class II · Ongoing

Class II FDA medical device recall by Getinge Disinfection Ab, initiated 2025-04-02. FDA's recall record names no specific 510(k) clearance for this event. Product: Washer Disinfector Aquadis 56; Model Number: 56M. Reason: The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing..

Recalling firm
Getinge Disinfection Ab
Classification
Class II
Status
Ongoing
Initiated
2025-04-02
Firm location
Vaxjo, N/A, Sweden

Product description

Washer Disinfector Aquadis 56; Model Number: 56M

Reason for recall

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Data sourced from openFDA. This site is unofficial and independent of the FDA.