Recall Z-1719-2018— Datascope Corporation
Class II · Terminated
Class II FDA medical device recall by Datascope Corporation, initiated 2017-10-19, terminated 2020-04-24. It names one FDA 510(k) clearance: K080592. Product: BEQ-TOP 38900 ECC PACK, Custom Tubing Kit, Catalog No. 701056368. Reason: The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow..
- Recalling firm
- Datascope Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-10-19
- Terminated
- 2020-04-24
- Firm location
- Fairfield, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BEQ-TOP 38900 ECC PACK, Custom Tubing Kit, Catalog No. 701056368
Reason for recall
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Data sourced from openFDA. This site is unofficial and independent of the FDA.