Recall Z-1719-2014— ConMed Corporation

Class II · Terminated

Class II FDA medical device recall by ConMed Corporation, initiated 2014-05-08, terminated 2017-04-11. It names one FDA 510(k) clearance: K120476. Product: MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical procedures to provide a path for RF current to leave the patient and return to the generator.. Reason: ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed..

Recalling firm
ConMed Corporation
Classification
Class II
Status
Terminated
Initiated
2014-05-08
Terminated
2017-04-11
Firm location
Utica, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical procedures to provide a path for RF current to leave the patient and return to the generator.

Reason for recall

ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed.