Recall Z-1714-2014— Toshiba American Medical Systems Inc

Class II · Terminated

Class II FDA medical device recall by Toshiba American Medical Systems Inc, initiated 2014-05-12, terminated 2014-10-02. It names one FDA 510(k) clearance: K120073. Product: Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.. Reason: Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop..

Recalling firm
Toshiba American Medical Systems Inc
Classification
Class II
Status
Terminated
Initiated
2014-05-12
Terminated
2014-10-02
Firm location
Tustin, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.

Reason for recall

Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.