Recall Z-1714-2014— Toshiba American Medical Systems Inc
Class II · Terminated
Class II FDA medical device recall by Toshiba American Medical Systems Inc, initiated 2014-05-12, terminated 2014-10-02. It names one FDA 510(k) clearance: K120073. Product: Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.. Reason: Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop..
- Recalling firm
- Toshiba American Medical Systems Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-05-12
- Terminated
- 2014-10-02
- Firm location
- Tustin, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.
Reason for recall
Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.