Recall Z-1713-2022— Cook Incorporated
Class II · Ongoing
Class II FDA medical device recall by Cook Incorporated, initiated 2022-07-19. It names one FDA 510(k) clearance: K110009. Product: Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-150 ORDER NUMBER (GPN): G56173. For vascular catheter positioning and exchange in diagnostic and interventional procedures.. Reason: Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised..
- Recalling firm
- Cook Incorporated
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-07-19
- Firm location
- Bloomington, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-150 ORDER NUMBER (GPN): G56173. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Reason for recall
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Data sourced from openFDA. This site is unofficial and independent of the FDA.