Recall Z-1710-2016— Synthes (USA) Products LLC

Class II · Terminated

Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2016-04-18, terminated 2017-02-02. FDA's recall record names no specific 510(k) clearance for this event. Product: 3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile Product Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).. Reason: The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier..

Recalling firm
Synthes (USA) Products LLC
Classification
Class II
Status
Terminated
Initiated
2016-04-18
Terminated
2017-02-02
Firm location
West Chester, PA, United States

Product description

3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile Product Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).

Reason for recall

The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.

Data sourced from openFDA. This site is unofficial and independent of the FDA.