Recall Z-1700-2020— Abbott

Class II · Terminated

Class II FDA medical device recall by Abbott, initiated 2020-02-28, terminated 2024-04-18. FDA's recall record names no specific 510(k) clearance for this event. Product: Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.. Reason: Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained..

Recalling firm
Abbott
Classification
Class II
Status
Terminated
Initiated
2020-02-28
Terminated
2024-04-18
Firm location
Sylmar, CA, United States

Product description

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.

Reason for recall

Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.