Recall Z-1699-2020— Diagnostica Stago, Inc.

Class II · Terminated

Class II FDA medical device recall by Diagnostica Stago, Inc., initiated 2020-03-04, terminated 2021-11-10. It names one FDA 510(k) clearance: K892859. Product: STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979. Reason: A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased.

Recalling firm
Diagnostica Stago, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-03-04
Terminated
2021-11-10
Firm location
Parsippany, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979

Reason for recall

A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased

Data sourced from openFDA. This site is unofficial and independent of the FDA.