Recall Z-1699-2015— Zimmer, Inc.
Class I · Terminated
Class I FDA medical device recall by Zimmer, Inc., initiated 2015-05-18, terminated 2016-08-02. It names one FDA 510(k) clearance: K071856. Product: M/L Taper with Kinectiv¿ Technology. prosthesis, hip, semi-constrained (metal uncemented acetabular component) Product Usage: Usage: Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis; disability due to previous fusion; previously failed endoprostheses, and/or total hip components in the affected extremity and acute femoral neck fractures.. Reason: Zimmer is initiating a voluntary recall of 64 lots (752 implants total) of M/L Taper with Kinectiv¿ femoral stems and modular necks due higher than allowed cytotoxicity levels found with the product. Reasonable probability of adverse biological response and subsequent revision.
- Recalling firm
- Zimmer, Inc.
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2015-05-18
- Terminated
- 2016-08-02
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
M/L Taper with Kinectiv¿ Technology. prosthesis, hip, semi-constrained (metal uncemented acetabular component) Product Usage: Usage: Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis; disability due to previous fusion; previously failed endoprostheses, and/or total hip components in the affected extremity and acute femoral neck fractures.
Reason for recall
Zimmer is initiating a voluntary recall of 64 lots (752 implants total) of M/L Taper with Kinectiv¿ femoral stems and modular necks due higher than allowed cytotoxicity levels found with the product. Reasonable probability of adverse biological response and subsequent revision
Data sourced from openFDA. This site is unofficial and independent of the FDA.