Recall Z-1698-2025— Vyaire Medical
Class I · Ongoing
Class I FDA medical device recall by Vyaire Medical, initiated 2025-04-10. It names one FDA 510(k) clearance: K151303. Product: Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber Component: No. a conduit for respiratory gas between the patient and a ventilator.. Reason: Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure..
- Recalling firm
- Vyaire Medical
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2025-04-10
- Firm location
- Mettawa, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber Component: No. a conduit for respiratory gas between the patient and a ventilator.
Reason for recall
Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.