Recall Z-1697-2024— Covidien, LP

Class II · Ongoing

Class II FDA medical device recall by Covidien, LP, initiated 2024-01-25. It names one FDA 510(k) clearance: K981128. Product: Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of natural silk fibers, for use in soft tissue approximation and/or ligation.. Reason: One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery..

Recalling firm
Covidien, LP
Classification
Class II
Status
Ongoing
Initiated
2024-01-25
Firm location
North Haven, CT, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of natural silk fibers, for use in soft tissue approximation and/or ligation.

Reason for recall

One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.