Recall Z-1696-2024— Advanced Instruments, LLC

Class II · Ongoing

Class II FDA medical device recall by Advanced Instruments, LLC, initiated 2024-03-11. FDA's recall record names no specific 510(k) clearance for this event. Product: OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0. Reason: System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls..

Recalling firm
Advanced Instruments, LLC
Classification
Class II
Status
Ongoing
Initiated
2024-03-11
Firm location
Norwood, MA, United States

Product description

OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0

Reason for recall

System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.