Recall Z-1696-2019— Accuray Incorporated
Class II · Terminated
Class II FDA medical device recall by Accuray Incorporated, initiated 2018-04-10, terminated 2021-04-01. It names one FDA 510(k) clearance: K150873. Product: CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS. Reason: There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach..
- Recalling firm
- Accuray Incorporated
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-04-10
- Terminated
- 2021-04-01
- Firm location
- Madison, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS
Reason for recall
There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.