Recall Z-1695-2026— Olympus Corporation of the Americas
Class II · Ongoing
Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2026-02-27. It names one FDA 510(k) clearance: K242191. Product: Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No. Reason: Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing..
- Recalling firm
- Olympus Corporation of the Americas
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-02-27
- Firm location
- Center Valley, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Reason for recall
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Data sourced from openFDA. This site is unofficial and independent of the FDA.