Recall Z-1691-2020— Microbiologics Inc
Class II · Terminated
Class II FDA medical device recall by Microbiologics Inc, initiated 2020-02-12, terminated 2021-03-26. It names one FDA 510(k) clearance: K861022. Product: Microbiologics QC SETS AND PANELS, YST Comprehensive QC Set, REF 5195P, containing Trichosporon dermatis, ATCC 204094, QR Code: (01)70845357030770, (17)200831, (10)5195-04.. Reason: KWIKSTIK 2 Pack (REF 0778P) and QC Set and Panels: KWIKSTIK YST Comprehensive QC Set (REF 5195P) contain product labeled as Trichosporon dermatis (T. dermatis) which may contain Candida Utilis instead of T. dermatis..
- Recalling firm
- Microbiologics Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-02-12
- Terminated
- 2021-03-26
- Firm location
- Saint Cloud, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Microbiologics QC SETS AND PANELS, YST Comprehensive QC Set, REF 5195P, containing Trichosporon dermatis, ATCC 204094, QR Code: (01)70845357030770, (17)200831, (10)5195-04.
Reason for recall
KWIKSTIK 2 Pack (REF 0778P) and QC Set and Panels: KWIKSTIK YST Comprehensive QC Set (REF 5195P) contain product labeled as Trichosporon dermatis (T. dermatis) which may contain Candida Utilis instead of T. dermatis.
Data sourced from openFDA. This site is unofficial and independent of the FDA.